Emerald System
SPECIFICATIONS
| Channel Configuration | 64/128 PR |
| Maximum applied voltage (50 Ω load) | PA: 150Vpp (Bipolar) / 75V (Unipolar) UT: 200V |
| Maximum PRF | ≤ 30 kHz |
| Acquisition | A-Scan/Peak/Conditional data recording |
| Acquisition triggered on | Free running, encoder position, external signal |
| Digitizing range | 800% |
| Max data file size | Limited by hard drive |
| Maximum number of reconstruction channels | 128 |
| Firing modes | FMC, PWI, Sparse* |
| TFM frame size (on board) | 1M points |
| TFM frame size (offline) | Unlimited |
| Simultaneous FMC channels | 2 |
| Max simultaneous TFM frames (on board) | 2 (up to 1M points per frame); 8 (up to 256k points per frame) |
| PA connector | 1 IPEX type (2 IPEX with splitter module) |
| UT connectors | 4 x LEMO® 00 (8 additional LEMO® 00 with splitter module) |
| Data connectivity | Ethernet 5 GBit/second |
| Encoder interfaces | 3 quadrature-type |
| I/O capability | 12 inputs, 9 outputs |
| Automated probe detection | Yes (with Zetec probe ID chip) |
| Automated scanner detection | Yes (with Zetec probe ID chip) |
| Gain | Up to 124 dB (0.1 dB step), 76 dB Analog / 48 dB Digital |
| Input impedance | 50 Ω |
| Bandwidth (-3 dB) | PA: 0.5 to 18 MHz, UT: 0.5 to 22.5 MHz |
| Data compression | 1, 2, 4, 8, 16 |
| Amplitude resolution | 14-bit elementary A-Scan, 16-bit PA |
| Max number of samples | 16k |
| Max A-scan range | 65k samples |
| Measurement gates | 6 + 1 synchronization gate (peak, crossing, autocrossing, homing) |
| Parallel PA channels processing | Up to 2 channels * |
| Rectification | Digital |
| Filtering | Analog / Digital (FIR) |
| Digitizing frequency | 100 MHz |
| TCG dynamic range | 40 dB |
Specifications listed may change without prior notification.
* – available in a future release