What's New:

|
Leading the pack.

Zetec’s MIZ-80iD combines the industry’s leading eddy current instrument, probe driver, power supply and MRPC® controller into one modular solution. To find out more,
click here.
|
|

|
|


|

|


|
MS 5800 Remote Data Acquisition System
The Zetec MultiScan MS 5800 is a portable multitechnology NDE (nondestructive evaluation) instrument that is suitable for a number of applications, such as: tube inspection, weld inspection, and corrosion mapping.
The system may combine any of the following NDT technologies: eddy current testing (ECT), remote field testing (RFT), magnetic flux leakage (MFL), and ultrasonic testing (UT). Each one of the technologies offered provides unique capabilities and superior performance.
|
The MS 5800 may also be used with probe pushers and scanners for automated inspections. The two built-in encoder inputs record the probe position, for accurate location of detected flaws. Acquisition and analysis are both performed from a standard computer using Zetec's EV or Zetec MultiView software. Both software packages allow for high volume data storage on any PC-compatible media.
Key Features:
Unique architecture expandable at any time to add ECT, RFT, MFL, rotating UT probes (IRIS), and multichannel UT options (up to eight channels available).
Communication with the computer is done through an Ethernet cable.
Lightweight: 12 kg, fully loaded.
High speed ECT inspections: 2 meters per second.
Very high immunity to ambient field interference. Some applications may be operated with 60 m (200 ft) probe cable extensions.
Compatible with ECT, RFT, and UT probes from other manufacturers.
Optional software (CARTO) for inspection planning and report generation on tube sheet diagrams.
Zetec EV and Zetec MultiView data acquisition and analysis software packages run on standard computers, with or without the MS 5800.
Three coil-drive modes: non-multiplex, multiplex, and super-multiplex.
Recording on any Microsoft® Windows®-supported storage devices.
Specifications

Power requirements:
|
Voltage: 100 - 240 VAC, ±10%
Fuse: 250 V, 2 A, slow-blow
Voltage: 100 - 240 VAC, ±10%
Maximum input current: 1.5 A
|
Environmental conditions:
|
Operating temperature: 0°C to 45° C (32°F to 113°F)
Storage temperature: -20°C to 60°C (-4°F to 140°F)
Relative humidity: 95%, no condensation
Safety: EN61010-1
|
MS 5800 housing:
|
Size (W x H x D): 465.0 mm x 380.0 mm x 255.0 mm (18.3 x 15.0 x 10.0 inches)
Net weight: 12.8 kg (28.2 lb)
Protection: Static proof case
|
Ethernet cable:
|
Type: Category 5, shielded
Maximum length: 100 m (328 ft)
|
Encoders:
|
Number of encoder axes: 2
Encoder type: Quadrature or Clk/Direction
Quadrature resolution: 4 counts/cycle
Frequency adjustment resolution: <0.1%
Input type: TTL
Counter size: 32 bits (from -2,147,483,648 to +2,147,483,648 counts)
Maximum frequency input: 200 kHz
Immunity to noise: Programmable filter from 8 Hz to 250 kHz
Sample rate control: Programmable axis selection and resolution
Input protection: Over voltage protection and ±15 V continuous
|
Auxiliary Inputs:
|
Type: Index, reset encoder, external clock, enable acquisition, and spare input
Index input: Used for C-scan index or rotation synchronization
Preset encoder: Used to preset encoder counter; programmable selection of preset counter
External clock: Used to control sample rate by external signal
Enable acquisition: Used to start/stop acquisition
Input type: TTL
Input protection: Overvoltage protection and ±15 V continuous
|
Auxiliary Outputs:
|
Type: External clock, relay outputs, and spare output
External clock: Used to synchronize the sample rate of other equipment with the MS 5800
Relay outputs: Spare outputs: Defined by the synchronization mode
|
Alarms:
|
Type of data: Eddy current, remote field, magnetic flux leakage, or ultrasound raw data
Setting of alarms: Programmable
Maximum response time: Smaller than the sampling period
Number of alarm outputs: 8 for electromagnetic techniques / 3 for ultrasound techniques
|
Options:
|
Eddy Current (ECT), Remote Field (RFT), Magnetic Flux Leakage (MFL), and Ultrasound (UT)
|
Purchase includes:
MultiScan MS 5800 User's Manual (French or English)
120 Vac and 240 Vac Power Cords
Ethernet Cable for PC connection
Carrying Case with wheels
Calibration certificates
Additional Information & Purchase:
MS 5800 ECT Data Sheet
MS 5800 ECT Array Data Sheet
MS 5800 R MFL Data Sheet
MS 5800 U Corrosion Mapping Data Sheet
MS 5800 U IRIS Probe Data Sheet
MS 5800 U Weld Inspection Data Sheet
Email us for current pricing and ordering information.
<< Full product listing
|

|
|


|